What does Statim G4 Cycle Fault 4 Mean?
The cassette has failed to achieve sterilization conditions within a time-out period of the chamber first reaching 110°C.
Probable Causes of Statim G4 Cycle Fault 4
1. Cassette Not Sealing Properly.
-To Repair This: Check for visible steam leaks from the cassette, if the cassette is leaking repair as needed. Replace cassette seal, lid or tray. -This is an End-user repair
2. Dirty Solenoid.
-To Repair This: Check the solenoid for debris and make sure the plunger is not sticking. - This is a Field Service Tech repair